EnvelopePower
SCPI Command :
CONFigure:WLAN:MEASurement<Instance>:MEValuation:LIST:SEGMent<segment>:ENPower
- class EnvelopePowerCls[source]
EnvelopePower commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get(segmentB=SegmentB.Default) float[source]
# SCPI: CONFigure:WLAN:MEASurement<Instance>:MEValuation:LIST:SEGMent<segment>:ENPower value: float = driver.configure.wlanMeas.multiEval.listPy.segment.envelopePower.get(segmentB = repcap.SegmentB.Default)
Specifies the expected nominal power of the measured RF signal for segment <no> in list mode.
- param segmentB:
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Segment’)
- return:
level: The range of the expected nominal power can be calculated as follows: Range (Expected Nominal Power) = Range (Input Power) + External Attenuation - User Margin
- set(level: float, segmentB=SegmentB.Default) None[source]
# SCPI: CONFigure:WLAN:MEASurement<Instance>:MEValuation:LIST:SEGMent<segment>:ENPower driver.configure.wlanMeas.multiEval.listPy.segment.envelopePower.set(level = 1.0, segmentB = repcap.SegmentB.Default)
Specifies the expected nominal power of the measured RF signal for segment <no> in list mode.
- param level:
The range of the expected nominal power can be calculated as follows: Range (Expected Nominal Power) = Range (Input Power) + External Attenuation - User Margin
- param segmentB:
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Segment’)